High Potency Glovebox Workstation System is designed as a complete process system with 2 Glovebox Workstations connected for processing and drying of HPAPI (High Potency Active Pharmaceutical Ingredients).
The system features include polypropylene superstructures, central transfer port, inlet HEPAs, dished black phenolic bases, top mount fans with BIBOs and HEPAs, independent hinged doors (24.635 W x 27” H opening), (8x) 10” oval glove ports, stainless steel tables, and glass viewing panels. Glass viewing panels and LED lighting maximize lighting across the workspace.Clean in Place CIP Wand and hose installed for misting inside of the enclosure.
Size:
168” Exterior Width
43” Exterior Depth
41” Interior Height
PERFORMANCE TESTING
Performance is paramount in the safety industry. Through consistent design quality and expert manufacturing, Flow Sciences’ units perform. With surrogate powder testing both in our facility as factory acceptance testing (FAT) and at the customer facility as site acceptance testing (SAT), we consistently exceed our customers’ containment targets and goal expectations.
FLOW SCIENCES CONTAINS. FLOW SCIENCES PERFORMS.
Containment Target : 50 ng/m³
Surrogate Powder Testing Result : 0.4 ng/m³
Equipment : Balances
Operation : Weighing / Transferring
Test : Naproxen Sodium – 3 x 500g