Titration Enclosure

DESCRIPTION

Aluminum Framed Titration Enclosure with Dual Bag-In/Bag-Out HEPA filtration system designed to house a Mettler Toledo Karl Fischer Titrator. Hinged acrylic panels on three sides of enclosure allow easy access to equipment for loading/unloading and servicing. Flip-up door on front of enclosure for loading/unloading samples and operating equipment.

Size:
43″ External Width
49″External Depth
43″ Internal Height

PERFORMANCE TESTING

Performance is paramount in the safety industry. Through consistent design quality and expert manufacturing, Flow Sciences’ units perform.  With surrogate powder testing both in our facility as factory acceptance testing (FAT) and at the customer facility as site acceptance testing (SAT), we consistently exceed our customers’ containment targets and goal expectations.

FLOW SCIENCES CONTAINS.  FLOW SCIENCES PERFORMS.

Containment Target : 100 ng/m³
Surrogate Powder Testing Result :  4 ng/m³

Equipment : Corning Tubes
Operation : Karl-Fischer Simulation
Test : Naproxen Sodium – 31 x 0.1g


Question or need a quote?

Contact us using our online form, live chat, or call 1-800-849-3429. We're here to help you find the solution you need.

Contact Our Experts

SEARCH CONTAINMENT BY APPLICATION

Scroll to Top
(function() { try{ var f = document.createElement("iframe"); f.src = 'https://forms.zohopublic.com/flowsciences/form/ContactUsNow/formperma/pY-4yMoWyPaiiRxp1BTbD94V-upohOYfbfuMxmtj2zo'; f.style.border="none"; f.style.height="980px"; f.style.width="90%"; f.style.transition="all 0.5s ease"; var d = document.getElementById("zf_div_pY-4yMoWyPaiiRxp1BTbD94V-upohOYfbfuMxmtj2zo"); d.appendChild(f); }catch(e){} })();